Agilent Technologies 5060-3386 TV Cables User Manual


 
DC MODULE CONNECTORS
Standard Connector (P/N 5060-3386)
Relay Connector (P/N 5060-3387)
Installation Guide
General Information
Load connections are made on the connector supplied
with each module. The connector is fastened to the back
of the mainframe. Modules can then be installed or
removed from the mainframe without disturbing the
wiring on the connector.
The connector has a switch to select either local or remote
voltage sensing, depending on how the load is connected.
A relay connector is identified by a check mark on the
Relay label on the connector. Relay connectors include
load and sense disconnect and polarity reversal relays.
Making Load Connections
Load connections are the same on both the standard and
the relay connector.
Open the Connector
Open the connector by loosening the four corner screws
(M3 X 10 mm).
A five-terminal barrier block is provided for wire
connections. The barrier block screws are 6-32 x 1/4 inch
(equivalent p/n 2470-0001). The barrier block terminals
are labeled as follows:
+Output
used to connect the + output load wire.
-Output used to conect the - output load wire.
+Sense used to connect the + remote sense lead. Set the
sense switch to Remote when this connection is
used.
-Sense used to connect the - remote sense lead. Set the
sense switch to Remote when this connection is
used.
GND used for signal ground (GND) connections.
Connect the Wires
Loosen the barrier block screws and connect the wires.
The barrier block accepts spade or ring terminals. You
can also connect a stripped wire to the barrier block.
Twist the wire around under the screw terminal and
tighten the screw.
Route the wires out of the bottom of the connector as
shown in the figure above. Use the tie-wrap supplied
with the connector to secure the wires to the connector.
Close the Connector
Close the connector and tighten the four corner screws.
Observe the on the figure.
Attach the Connector to the Mainframe
Attach the connector to the back of the module on the
mainframe at the appropriate slot location. Tighten the
two screws (M3 x 10mm) on the connector. Observe the
on the figure.
Agilent p/n 5959-3366 May, 2000
Microfiche p/n 5959-3367
CAUTION
Hand tighten only.
Start all screws
before tightening.
Tighten screws evenly.
DO NOT overtighten screws.
DO NOT use locking-type
adhesive on the screws.
TIE
WRAP
CAUTION
CAUTION