HP (Hewlett-Packard) 86140A TV Cables User Manual


 
4-59
Remote Operation
CALCulate Subsystem Commands
CALCulate:MARKer[1|2|3|4]:X:TIME <numeric_value>
Sets the X-axis value of the normal marker when the instrument is in zero
span. When the delta function is
ON
, the absolute X-axis value of the delta
marker is controlled. When the bandwidth function is
ON
, the X-axis value of
the center marker is controlled. The default units of the parameter is seconds.
Sending the command when the specified marker is
OFF
will turn the marker
ON
and place the marker at the desired position. Sending the command when
the instrument is in a non-zero span will generate a Settings conflict error.
CALCulate:MARKer[1|2|3|4]:X[:WAVelength] <numeric_value>
Sets the X-axis value of the normal marker in wavelength units. When the
delta function is
ON
, the absolute X-axis value of the delta marker is con-
trolled. When the bandwidth function is
ON
, the X-axis value of the center
marker is controlled. The default units of the parameter are meters.
Sending the command when the specified marker is
OFF
will turn the marker
ON
and place the marker at the desired position. Sending the command when
the instrument is in a zero span will generate a Settings conflict error.
CALCulate:MARKer[1|2|3|4]:Y?
Returns the Y-axis value of the normal marker. When the delta function is
ON
,
the value returned is the absolute Y-axis value of the delta marker. When the
bandwidth function is
ON
, the value returned is the Y-axis value of the center
marker.
Sending the command when the specified marker is off will generate a Set-
tings conflict error.
book.book Page 59 Monday, January 31, 2000 10:34 AM