HP (Hewlett-Packard) 86140A TV Cables User Manual


 
4-85
Remote Operation
SENSe Subsystem Commands
and
[SENSe:][WAVelength:]CENTer:STEP:AUTO OFF|ON|0|1
[SENSe:][WAVelength:]CENTer:STEP:AUTO?
When
ON
, the step size is automatic. When
OFF
, the step size is fixed. The
value is set by the [SENSe:][WAVelength:]CENTer:STEP[:INCRement] com-
mand.
[SENSe:][WAVelength:]CENTer:STEP[:INCRement] <numeric_value>[M|NM|UM|A]
[SENSe:][WAVelength:]CENTer:STEP[:INCRement]?
Specifies the center wavelength step size.
[SENSe:][WAVelength:]LIMit OFF|ON|0|1
[SENSe:][WAVelength:]LIMit?
Specifies whether the span is limited to the specified range of 600 to 1700 nm.
[SENSe:][WAVelength:]OFFSet <numeric_value>[M|NM|UM|A]
[SENSe:][WAVelength:]OFFSet?
Specifies the wavelength offset. This is the offset between the measured
wavelength and the displayed wavelength.
Center Stop
Span
2
--------------""""=
With Wavelength Limit Off, the minimum value for the Center Wavelength is nominally
350.1 nm. The maximum value is 1999.9 nm. These limits are valid for wavelengths ref-
erenced in air or vacuum.
With Wavelength Limit On, the minimum value for the Center Wavelength is nominally
600.1 nm. The maximum value is 1699.9 nm. These limits are valid for wavelengths ref-
erenced in air or vacuum.
The Preset value for Wavelength Limit is On. The Preset value for Center Wavelength is
1150 nm.
book.book Page 85 Monday, January 31, 2000 10:34 AM