SENSe Subsystem Commands
[SENSe:][WAVelength:]STOP <numeric_value>[M|NM|UM|A|HZ|KHZ|MHZ|GHZ]
[SENSe:][WAVelength:]STOP?
Specifies the stop wavelength. The center wavelength and span are adjusted
so that:
and
If the instrument is in zero span, this command sets the center wavelength to
the value specified.
Start Center
Span
2
--------------=
Stop Center
Span
2
--------------+=
With Wavelength Limit Off, the minimum value for the Stop Wavelength is nominally
350.2 nm. The maximum value is 2000 nm. These limits are valid for wavelengths refer-
enced in air or vacuum.
With Wavelength Limit On, the minimum value for the Stop Wavelength is nominally
600.2 nm. The maximum value is 1700 nm. These limits are valid for wavelengths refer-
enced in air or vacuum.
The Preset value for Wavelength Limit is On. The Preset value for Stop Wavelength is
1700 nm.