Printer Head Cleaning Procedure
Figure 6-1. Example of a static-safe workstation
Procedure
1
Turn off the HP 86140-series optical spectrum analyzer, and remove the line
power cord.
2
Place the instrument at a static-safe work station as described in the
introduction to this procedure.
3
Use a coin or screwdriver to open the printer door that is located on the top of
the instrument.
CAUTION
Avoid dropping the coin or screwdriver, used to open the printer door, into the
printer assembly.
4
Lift up the paper latch as shown in the following diagram, and remove the
paper.
book.book Page 8 Monday, January 31, 2000 10:34 AM