HP (Hewlett-Packard) 86140A TV Cables User Manual


 
3-55
Using the Optical Spectrum Analyzer
The System Menus
To set the HP-IB address
1
Press the front-panel
System
key.
2
Press the
More System Functions....
softkey.
3
Press the
Remote Setup....
softkey. The Remote Setup panel appears.
4
The Current HPIB Address information shows the HP-IB address currently
being used. To change the address, use the numerical entry pad or knob to
change the New HPIB Address information. Press the
Defaults
softkey to reset
the address to factory preset defaults. Refer to To Fill In a Setup Panel on
page 3-65 for information on changing and selecting items in the setup panel.
5
Press the
Set HPIB Address
softkey if changes were made.
The Remote Setup panel
Setup panel
selections
Current HPIB Address
Shows the HP-IB address currently being used.
New HPIB Address
Enter the new HP-IB address in the text box using the number keys or knob.
Note
None of the Automeasure Setup parameters are affected by the front-panel Preset key.
They are not saved as part of the measurement setup.
book.book Page 55 Monday, January 31, 2000 10:34 AM