Specifications and Approvals Teledyne API T703/T703U Calibrator Operation Manual
22
Table 2-3: T703/T703U Specifications for O
3
Photometer
Ranges 0-100 ppb to 0-10 ppm, user selectable
Zero Noise 0.3ppb (RMS)
Span Noise <0.5%
Lower Detectable Limits 0.6 ppb (RMS)
Precision 1.0 ppb
Linearity 1.0% of Full Scale
Lag Time <10 seconds
Rise/Fall Time <20 sec (photometer response)
Zero Drift <1.0 ppb / 7 days
Span Drift <1% / 24 hours; <2% / 7 days
Flow Rate
800 cc
3
/min +/- 10%
2.2. APPROVALS AND CERTIFICATIONS
The Teledyne API Models T703 and T703U Photometric O
3
Calibrators were tested and
certified for Safety and Electromagnetic Compatibility (EMC). This section presents the
compliance statements for those requirements and directives.
2.2.1. SAFETY
IEC 61010-1:2001, Safety requirements for electrical equipment for measurement,
control, and laboratory use.
CE: 2006/95/EC, Low-Voltage Directive
North American:
cNEMKO (Canada): CAN/CSA-C22.2 No. 61010-1-04
NEMKO-CCL (US): UL No. 61010-1 (2nd Edition)
2.2.2. EMC
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity
EN 55011 (CISPR 11), Group 1, Class A Emissions
FCC 47 CFR Part 15B, Class A Emissions
CE: 2004/108/EC, Electromagnetic Compatibility Directive
2.2.3. OTHER TYPE CERTIFICATIONS
For additional certifications, please contact Technical Support: