9.6.2 Calibrator Commands
DAMP
Turns the dampening function on or off.
For example:
If you send DAMP ON this will turn the dampening function
on.
DAMP?
Returns the current state of the dampening function.
For example:
If you send DAMP? It will return ON if the dampening
function is on.
DISPLAY
Turns the indicated display on or off.
For example:
If you send DISPLAY LOWER, ON this will turn the lower
display on.
DISPLAY?
Returns the current state of the each of the displays.
For example:
If you send DISPLAY? It will return ON, ON, ON if the all
the displays are on.
FAULT?
Returns the error code number of an error that has occurred.
The command may be entered when the previous command did
not do what it was meant to do.
For example, if a value for current output is entered that is bigger
than the supported range (0-24mA) FAULT? Would return:
103 which is the code number for an entry over range.
Refer to the Error Codes table for more information on error code
numbers.