7.0 PARAMETERS
The balance has 8 parameters that can be set by the user.
FUNCTION SECTION DESCRIPTION
F1 Unt
See section 7.1 Sets the units to be used
g, ct, GN, dwt, oZt, Oz, Lb, MM, t, tLH, tLS, tLT, tLJ
F2 HLd
See section 7.2 Sets the operation of the Hold function
On HD: Manual hold the weight whit the [Hold/Print] key
Auto HD: Weight held automatically when stable
OFF HD: Hold function Off, Density function disabled
F3 bL
See section 7.3 Sets the backlight
on BL
: backlight always on
AU BL
: backlight automatically turns on when a weight is
placed on the pan or a key is pressed
oFF BL:
backlight always off
F4 SEr
See section 7.4 Sets the print parameters
F5 OFF
See section 7.5 Sets the auto power-off parameter
F6 ICE
See section 7.6 Selects Internal or external calibration mode
F7 CA
See section 7.7 Calibration mass value fine adjustment
F8 SEt
See section 7.8 User Parameters for auto zero, filter and stability
tECH
Technical parameters setting mode / factory setting