6 - 4
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — System configuration and diagnostics
↓↓
Diagnostic menus
↓↓
Control module diagnostics
↓↓
In the menu "Control Unit Diagnostics" you can find software error messages of the
control module. If such messages are existing, you can act in the following way:
1) Make a note of the error messages.
2) Reset the message:
• Press the 88 -key or the →→ -key in the line "Edit to reset".
• Set up "Reset" with the ↑↑ -key or the ↓↓ -key and confirm it with the 88 -key:
The error messages will disappear, if their reason will not exist any longer.
• If the error messages will keep displayed:
3) System reset:
• Change to the menu "System Reset" with the F5 key (Reset...).
• Press the 88 -key or the →→ -key in the line "System reset !" to reboot (see 6.5 p. 6-11).
4) Check the error messages:
• Go again to the menu "Control Unit Diagnostics".
• If the error messages will still exist, contact your customer service, please.
• If the error messages will be disappeared, you should set up the report status again:
Set up "Report" in the line "Edit to reset".
Measure Back... Reset...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Control Unit Diagnostics --
Software error code (1 = no error): 1
And:
And:
And:
And:
And:
And:
Edit to reset: Report
Last software error message:
And:
6.1 Diagnostic Menus
6.1.1 Control Module Diagnostics