3 - 7
No. 99MBC071A
3. DISPLAYS AND KEY OPERATIONS
3.1.2.6 Measurement mode
This mode can be divided into the following operational states:
1) Measurement in the ready state
This is the measurement mode that is entered immediately after the power is turned on
or if another measurement mode is aborted by pressing the
C
key (or by the RESET
signal from the I/O interface or the “CL” command from the RS-232C/GP-IB interface).
It is used to establish setups for calibration and available functions, which are not part
of the basic setup items, or to enter another measurement mode including single-run
measurement.
Usually GO/NG judgment and analog output will not take place for measurement in the
ready state, however, these specifications can be made in the basic setup mode.
Measurements in the ready state are unavailable for statistical processing.
2) Single-run measurement
If the
RUN
key (otherwise input RUN via the I/O interface or “R” command via the
RS-232C/GP-IB interface) is pressed, one session of measurement is performed and the
results will be automatically subject to GO/NG judgment and analog output. In addition,
the measured data will be outputted for the RS-232C/GP-IB interface, Digimatic Output
Unit, and printer. The measured data will be held (latched for the specified period) in
the display.
This data will be available for statistical processing.
3) Continuous-run measurement
If the
C.RUN
key (otherwise input RUN+RESET via the I/O interface or “CR” command
via the RS-232C/GP-IB interface) is pressed, one session of measurement is started and
repeated the specified number of times. The measured data will be automatically subject
to GO/NG judgment and analog output. In addition, the measured data will be outputted
for the RS-232C/GP-IB interface, Digimatic Output Unit, and printer.
Press the
RUN
or
C.RUN
key (or if RUN is received from the I/O interface) again to
terminate the measurement and hold the measured data on the display. If the
C
key
(or input RESET via the I/O interface or “CL” command via the RS-232C/GP-IB interface)
is pressed halfway, the measurement is aborted and the ready state is returned to.
The measurements are available for statistical processing.
4) Continuous measurement with a term specification
This will take place where RUN input from the I/O interface has been assigned so as to
start a term-specified continuous-run measurement in the basic setup.
Repeatedly performs single-run measurement while RUN signal input continues, which
is basically the same as the continuous-run measurement. Therefore, hereafter, continu-
ous-run measurement includes the ones with a term specification.
The measurements are available for statistical processing.