3 - 2 Users Manual
The instrument described in this manual is to be used by properly trained
personnel.
Do not remove the cover or perform any adjustment, maintenance, or repair
unless you are qualified to do so and are aware of the hazards involved.
Symbol Before connectin
the instrument to line power, ensure that the
power
round is functionin
correctly. Before any other connection
is made, the instrument must be connected to the earth (
round)
conductor via the three-conductor power cord. The power plu
must be inserted into a socket outlet provided with
round contact.
The
roundin
protection must not be defeated by use of an
extension cord without a
round conductor.
WARNING : Any interruption of the protective conductor inside or outside
the instrument or disconnection of the protective
round
terminal is likely to make the instrument dan
erous.
Intentional interruption is prohibited.
The instrument is shipped from the factory with a power cord that has a plu
appropriate for the country in which it is sold. If a different power plu
or power
cord is required, contact your Fluke Sales and Service Center.
CAUTION: When an instrument is brou
ht from a cold into a warm environment,
condensation may cause a hazardous condition; therefore, make
sure that the
roundin
requirements are strictly adhered to.
3.4 INSTALLATION
3.4.1 Mechanical installation
The instrument is fan cooled and must be installed in a location that allows
sufficient space at the three important sides: rear, left, and ri
ht. Each side must
have a space of at least 1 inch (25 mm). The fan, located at the rear, cools the
instrument by drawin
air into the openin
s at the rear and exhaustin
it throu
h
openin
s on the sides. Each instrument output channel has its own fan.