8
CAUTIONS FOR USAGE
■ The top surface of the frame sometimes becomes hot during the initial operation or when operating in a
hot environment. This is due to the hot gas circuit for preventing dew, and is not an indication of a
malfunction.
■ This unit is provided with the filter check lamp. If the filter check lamp is flashed, clean the condenser
filter. Refer to page 25 “Cleaning of condenser filter”. A clogged filter may cause poor cooling and
compressor trouble.
■ When connecting a liquid CO
2
cylinder for the back-up system , consult your dealer of a neighboring
engineering shop of high pressure gas.
■ Leave the remote alarm switch “OFF” until the inside temperature of the freezer reaches the
appropriate level. Leave the back-up system switch “OFF” as well. After the inside temperature
reaches the appropriate level, set the needle for the warning device operation of the temperature
recorder.
■ When the unit is first installed, or re-started after long period of no use, the power failure alarm may not
operate, because its Cadnica batteries have completely discharged. This does not denote a malfunction.
In this case, continue to operate the unit for about 3 hours. To fully recharge the completely discharged
batteries, the unit has to be operated for about 2 days.
■ Cover the control panel with the control panel cover to avoid the malfunction resulting from an
accidental contact with control panel (such as power switch OFF).