Teledyne API Model 200AH NO
X
Analyzer Instruction Manual, 06492, Rev. G2
2-9
Step 2 - Calibrate the instrument for NO
X
:
Zero/Span Calibration Procedure
Step
For Units w/o
O
2
Option
For Units with
O
2
Option
Comment
1. Input NO
X
Zero gas Input NO
X
Zero gas Allow Zero gas to enter the sample port on
the rear of the instrument.
2. Press CAL Press CAL-NOX The M200AH enters the calibrate mode from
sample mode.
3. Wait 10 min Wait 10 min Wait for reading to stabilize at the zero value.
If you wait less than 10 minutes the final zero
value may drift.
4. Press ZERO Press ZERO The ZERO button will be displayed.
5. Press ENTR Press ENTR Pressing ENTR actually changes the
equations and zeros the instrument.
6. Press EXIT and
input NO
X
Span gas
Press EXIT and
input NO
X
Span gas
M200AH returns to the CAL menu. Now
switch gas streams to span gas.
7. Wait 10 min Wait 10 min Wait for reading to stabilize at the span
value.
8. Press SPAN Press SPAN The SPAN button should be displayed. If
there is no SPAN button check the
Troubleshooting Section 9.2.8 for
instructions on how to proceed. In certain
circumstances at low span gas concentrations
both the ZERO and SPAN buttons will
appear.
9. Press ENTR Press ENTR Pressing ENTR actually changes the
equations so that the concentration displayed
is the same as the expected span
concentration you entered above, thus
spanning the instrument.
10. Press EXIT Press EXIT Pressing EXIT returns the instrument to
SAMPLE mode.
11.
Input
O
2
Zero gas
Allow Zero gas to enter the sample port on
the rear of the instrument.
12. Press CAL-O2 The M200AH enters the calibrate mode from
sample mode.
(table continued)