If the no-load calibration value is acceptable, the display will then show
""
where


is the value of calibration mass required and “
""
is the units (kg or lb) for the calibration mass, e.g.

for a 6 kg
mass or “

” for a 10 pound mass.
Place the correct calibration mass on the scale.
Press the [Cal] key again. The display shows
” until calibration is
complete and then returns to normal weighing.
Verify the calibration by placing the calibration mass on the scale. The
display should show the correct calibration mass in the currently
selected units.
NOTE:
If the no-load or full-load reading is too low

will be displayed
If the no-load or full-load reading is too high

will be displayed
If the no-load or full-load reading is unstable
will be displayed.
If any of these conditions occurs, or if the [Cal] key is not pressed within 10
seconds of the “

” message being displayed, the calibration procedure will
be aborted and the scale will return to normal weighing using the previous
tare and calibration values.
6.0 USER SET UP PARAMETERS
6.1 FUNCTION OF THE KEYS
To enter the user-parameter setup menu, the [Zero/Tare] key should be held
pressed, and while the display is showing  the [Units] key should be
pressed. The first setup option will then be shown. Release both the keys. If
[Zero/Tare] is released without the [Units] key being pressed, the display
will change to
, a normal tare operation will be performed, and the
setup menu will not be entered.
When in the setup menu, the functions of the keys are as follows:
Press [Zero/Tare] to select alternatives within any parameter.