FREEZER COMPONENTS
1. Lock: Turn clockwise to 180
o
with a key and the outer door is securely locked.
2. Magnetic door gasket: This prevent the cold air from escaping. Always keep clean.
3. Handle: Always grip this handle to open and close the outer door.
4. Drain hose: The drain hose can be used to collect the defrosted water. See page 17 for usage.
5. Leveling foot: The height of the freezer can be adjusted by this screw type foot. Keep the unit in
level at the installation.
6. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling foot so that the front 2 casters cannot contact with the floor.
7. Drain cap: These are drain in defrosting and drain cap. See page 17 for usage.
8. Shelf: Items to be stored in the chamber must be placed on the shelves. The maximum storage
weight for each shelf is 50kg.
9. Access port: This is used for leading the measuring cable from the freezing chamber to the outside.
A total of two(2) are provided. ; left side and top of the freezer.
10. Door latch: The lock the door, turn this latch downward. A padlock is also available.
11. Temperature sensor: Located on the rear of the chamber.
12. Control panel: The operation status is displayed on this panel. And the temperature setting is
available through this panel. Refer to page 10 for details.
13. Front cover: Open this cover when connecting the remote alarm.
14. Space for temperature recorder: Space for an automatic temperature recorder (optional
component) is available separately. See page 25 “TEMPERATURE RECORDER (OPTION)”.
15. Battery switch: Switch for battery used for power failure alarm. Always keep ON. Turn the
switch OFF when the unit is in no use for a long period (more than 1 month).
16. Remote alarm terminal: Used to notify an alarm condition of the unit to remote location. See
page 14 for details.
17. Power switch (back side): This is for turning ON/OFF the power to the unit. ON – “I” OFF – “
18. Fixture (back side): These keep the adequate apace between the unit and wall, and also can be
used for fixing the unit.