8472B Operating and Service Manual 1
Description
Description
The 8472B crystal detector is a 50
Ω (
nominal) device designed for
measurement use in coaxial systems.
Features The instrument converts RF power levels applied to the 50
input
connector into proportional values of dc voltage.
The instrument measures relative power up to 200 mW and has a BNC
female connector for the output jack which allows the detected output to
be connected to a SWR meter.
The output voltage polarity is negative, unless Option 003 is selected.
The frequency range of the detector is 10 MHz to 18 GHz.
Options The 8472B crystal detector is available with the following options. Refer to
Table 1 for further descriptions.
Option 001: Matched pair of detectors
Option 002: Optimum square law load
Option 003: Positive polarity output
Option 100: Female OSSM type output connector
Mating Connectors The mating output connector used with the crystal detector must be a
male BNC connector for the standard output connector or an OSSM
male connector for Option 100.
The mating RF input connector must be a female SMA connector.
CAUTION SMA connectors have a limited life in applications that require repeated
connecting and disconnecting. In-series adapters should be used for such
applications.