Agilent Technologies ES TV Converter Box User Manual


 
Chapter 1 1-7
Installing Your Analyzer
STEP 3. Meet Electrical and Environmental Requirements
STEP 3. Meet Electrical and Environmental Requirements
1. Set the line-voltage selector to the position
that corresponds to the AC power source.
2. Ensure the available AC power source
meets the following requirements:
90–132 VAC
47–66 Hz / 400 Hz (single phase)
- or -
198–265 VAC
47–66 Hz (single phase)
The analyzer power consumption is 350 VA
maximum.
3. Ensure the operating environment meets
the following requirements:
4. Verify that the power cable is not damaged,
and that the power-source outlet provides a
protective earth contact.
0 to 55 °C
< 95% relative humidity at 40 °C
(non-condensing)
< 15,000 feet ( 4,500 meters) altitude
Some analyzer performance parameters are
specified for 25 °C ±5 °C. Refer to the
Reference Guide for information on the
environmental compatibility of warranted
performance.
WARNING Any interruption of the
protective (grounding)
conductor or disconnection
of the protective earth
terminal, can result in
personal injury, or may
damage the analyzer.