Fluke Fluke 16 Universal Remote User Manual


 
PN 644226 July, 1997 Rev.1, 3/99
© 1997,1999 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.
1
®
16
Multimeter with Temperature
Calibration Information
W
Warning
To avoid electrical shock, remove test leads and any input signals from the Meter before opening the case.
Caution
This Meter contains parts that can be damaged by static discharge. Follow the standard practices for handling
static sensitive devices.
Introduction
The 16 Calibration Information instruction sheet provides the
information necessary to calibrate and verify the Fluke
Model 16 Multimeter with Temperature (hereafter known as
the Meter). This instruction sheet provides the following
information:
Product specifications
Disassembly and reassembly
Performance test procedures
Calibration procedures and adjustments
Replacement parts
See the instruction sheet for operating instructions.
Service
The Meter is warranted to be free from defects in material
and workmanship for one year. The complete warranty
statement is in the Meter’s instruction sheet.
To contact Fluke, call one of the following telephone
numbers:
USA and Canada: 1-888-99-FLUKE (1-888-993-5853)
Europe: +31-402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-356-5500
Or, visit Fluke’s Web stie at www.fluke.com.
Specifications
Specifications are in Table 1. Accuracy is specified for a
period of one year after calibration, at 18ºC to 28ºC (64ºF to
82ºF) with relative humidity to 90%. AC conversions are ac-
coupled, average responding, and calibrated to the rms
value of a sine wave input.
Table 1. Specifications
Maximum Voltage Between any Terminal and Earth Ground (excludes 10% tolerance) 600 V ac or dc
Display 3-3/4-digits, 4000 counts, updates 4/sec
Operating Temperature -10°C to 50°C
Storage Temperature -30°C to 60°C indefinitely (to -40°C for 100 hrs)
Temperature Coefficient 0.1 x (specified accuracy)/°C (<18°C or >28°C)
Relative Humidity 0% to 90% (-10°C to 35°C)
0% to 70% (35°C to 50°C)
Altitude Operation/Non-operation 2000 m
Battery Type 9V, NEDA 1604 or IEC 6F22
Battery Life 650 continuous hours with alkaline
450 continuous hours with carbon-zinc
Shock, Vibration 3 meter drops: Sinusoidal vibration up to 3 G @ 55 Hz
Size (HxWxL) 1.35 in x 2.75 in x 5.55 in
(3.46 cm x 7.05 cm x 14.23 cm)
Weight 10 oz (286g)
EMI Regulations Complies with FCC Part 15B, Class B, EN50081-1, EN50082-1.
Safety Designed to Protection Class II requirement of UL1244, ANSI/ISA-S82.01 - 1988, CSA C22.2 No 231,
and VDE 0411, and IEC 1010-1 overvoltage category III (CAT III), 600 V
Certification
P $