16
Multimeter with Temperature
2
Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits]).
Table 1. Specifications (cont)
Function Range Resolution Accuracy
f
(50 to 400 Hz)
4000 mV
1
4.000 V
40.00 V
300.0 V
400.0 V
600.0 V
1 mV
0.001 V
0.01 V
0.1 V
0.1 V
1 V
±(1.9%+3)
±(1.9%+3)
±(1.9%+3)
NA
±(1.9%+3)
±(1.9%+3)
E
4000 mV
1
4.000 V
40.00 V
300.0 V
400.0 V
600 V
1 mV
0.001 V
0.01 V
0.1 V
0.1 V
1 V
±(0.9%+2)
±(0.9%+2)
±(0.9%+1)
NA
±(0.9%+1)
±(0.9%+1)
J
400.0Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
0.1Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
±(0.9%+2)
±(0.9%+1)
±(0.9%+1)
±(0.9%+1)
±(0.9%+1)
±(1.5%+3)
K
1.000 µF
10.00 µF
100.0 µF
1000 µF
10000 µF
0.001 µF
0.01 µF
0.1 µF
1 µF
10 µF
±(1.9%+2)
±(1.9%+2)
±(1.9%+2)
±(1.9%+2)
±(10%+90)
typical
µ[ 0 to 200 µA 0.1 µA ±(1% + 2 counts)
µ
\
(50 to 400 Hz)
0 to 200 µA 0.1 µA ±2% + 3 counts)
Temperature
3
(Type K Thermocouple)
-10°C to 400°C
14°F to 75°F
-40°C to -10°C
-40°F to 14°F
0.1°C
0.2°F
0.1°C
0.2°F
±(1% + 0.8°C)
±(1% + 1.5°F)
±(5% + 1.5°C)
Typical
±(5% + 3.3°F) Typical
C
2
2.000 V 0.001 V ±(0.9%+2)
1. The 4000 mV range can be entered only in the manual range mode. Use the 4000 mV range with accessories.
2. The beeper is guaranteed to come on at <25Ω and turn off at >250Ω. The Meter detects opens or shorts of 250 µs or
longer.
3. Error does not include Type K Thermocouple errors.