No labels specified. Indicates there are no labels to which to assign
symbols.
(x) Occurrences Remaining in Sequence (y). Indicates the logic analyzer
is waiting for (x) number of occurrences in sequence level (y) of the state
trace specification before it can go on to the next sequence level.
(x) Secs Remaining in Trace. Indicates the amount of time remaining
until acquisition is complete in Glitch mode.
Search failed - O pattern not found. Indicates the O pattern does not
exist in the acquired data. Check for a correct O marker pattern
specification.
Search failed - X pattern not found. Indicates the X pattern does not
exist in the acquired data. Check for a correct X marker pattern
specification.
Slow Clock or Waiting for Arm. Indicates the state analyzer is waiting
for a clock or arm signal. Re-check the state clock or arming specification.
Slow or missing Clock. Indicates the state analyzer has not recognized a
clock for 100 ms. Check for a missing clock if the intended clock is faster
than 100 ms. If clock is present but is slower than 100 ms, the data will
still be acquired when a clock is recognized and should be valid.
Specified inverse assembler not found. Indicates the inverse assembler
specified in the configuration file cannot be found on the disc.
State clock violates overdrive specification. Indicates the data from the
last acquisition is not reliable due to the state clock signal not being
reliable. Check the clock threshold for proper setting and the probes for
proper grounding.
(x) States Remaining to Post Store. Indicates the number of states
required until memory is filled and acquisition is complete.
Time correlation of data is not possible. "Count" must be set to "Time"
in both machines to properly correlate the data. This message is also
displayed when the data from this state/timing module cannot be time
correlated in an intermodule "Group Run" configuration.
Time from arm to trace point > 41.943 ms. The correlation counter
overflows when the time from a machine’s arm to the machine’s trigger
Error Messages HP 16510B
B-2 Front-Panel Reference