To set up the input coupling test
Performance Test only
This test verifies that the Agilent 35670A meets its input specification for ac
coupling rolloff. In this test, a frequency synthesizer outputs a 1 Hz signal to all
channels. The signal is measured in both ac and dc coupled modes. The value
measured in ac coupled mode is subtracted from the value measured in dc coupled
mode to determine the ac coupling rolloff.
1 2ch
1 4ch
Verifying Specifications Agilent 35670A
To set up the input coupling test