To set up the single channel phase accuracy test
Performance Test and Operation Verification
This test verifies that the Agilent 35670A meets its single channel phase accuracy
specification. In this test, a frequency synthesizer outputs an identical square wave
to all channels and a synchronized TTL-level signal to the trigger input. The phase
difference between the trigger and each channel is measured to determine phase
accuracy.