Compur 4120 VCR User Manual


 
Compur Statox 4120
Table 4: Switch position for Cl
2
Measuring range: 0-1,5 ppm 0-3 ppm 0-10 ppm
S1/1 ON ON OFF
S1/2 ON ON ON
S1/3 ON ON ON
S1/4 ON OFF ON
S1/5 ON ON OFF
S1/6 ON ON ON
S1/7 OFF OFF OFF
S1/8 ON ON ON
E-PROM Index sensor head 12, 13, 14 12, 13, 14 12, 13, 14
E-PROM Index control module 02, 03 02 02, 03
Table 5: Switch position for CO
Measuring range: 0-150 ppm
S1/1 ON
S1/2 ON
S1/3 ON
S1/4 ON
S1/5 ON
S1/6 OFF
S1/7 OFF
S1/8 ON
E-PROM Index sensor head 12, 13, 14
E-PROM Index control module 02
Table 6: Switch position for HCl Table 7: Switch position for ClO
2
Measuring range: 0-100 ppm Measuring range: 0-0,5 ppm
S1/1 OFF S1/1 OFF
S1/2 ON S1/2 ON
S1/3 ON S1/3 ON
S1/4 ON S1/4 OFF
S1/5 ON S1/5 ON
S1/6 ON S1/6 ON
S1/7 ON S1/7 ON
S1/8 ON S1/8 ON
E-PROM Index sensor head 19 E-PROM Index sensor head 17
E-PROM Index control module
02
E-PROM Index control module
02