2233
QUICK-START
1. On the Remote, press PROG, NEXT, ENTER.
2. Use the arrow buttons to select COnduc (conductivity),
nAOH (Sodium Hydroxide 0-15%), HCL (Hydrochloric
Acid 0-16%), H2SO4L (Sulfuric Acid 0-30%), H2SO4H
(Sulfuric Acid 95-99.99%), or CuSt (custom curve)
mode. Press ENTER. If you chose CuSt, continue with
step 3. If you chose COnduc or one of the prepro-
grammed % concentration modes, skip step 3 and go to
step 4.
3. If you selected CuST, you will see the Setup Custom
screen. To move to the custom curve configuration
menu, press ENTER. You will automatically return to
this same Setup Custom screen after configuration is
complete. To continue transmitter display programming,
press NEXT in the Setup Custom screen.
4. Use the arrow keys to toggle temperature units between
Celsius and Farenheit.
5. Press ENTER then RESET.
6. Press PROG, ENTER.
7. Use the arrow key to toggle t AutO to On or OFF to
select using either the process temperature (tAutO =
On) or a manual temperature (tAutO = OFF). Press
ENTER. If you selected t AutO = OFF, you will be
prompted to enter the manual temperature; use the
arrow keys, then press ENTER.
8. If you selected CondUC in step 2, you will see a COMP
(Temperature Compensation type) screen. Use the
arrow keys to select desired temperature compensa-
tion: LinEAr (linear) or nOnE (raw or uncompensated
conductivity). Press ENTER. If you are in LinEAR
mode, you can now enter a particular temperature slope
(default is 2%/degC), then press ENTER to apply the
slope.
9. Press RESET.
10. Press CAL, NEXT, NEXT, NEXT, ENTER.
11. Use the arrow buttons to enter the cell constant of the
sensor. Press ENTER, then EXIT.
12. To “zero” the sensor in air, press CAL, NEXT, ENTER.
13. Hold the sensor in air to zero. Press ENTER, then EXIT.
14. If you are measuring % concentration (nAOH, HCL,
H2SO4L, or H2SO4H) or custom curve (CuSt), quick
start is complete; proceed to step 20.
15. If you are measuring conductivity (CondUC), then stan-
dardize the sensor by placing the sensor in a solution
of known conductivity value. Press CAL, ENTER.
16. Use the arrow buttons to enter the current conductivity
value of the solution. Press ENTER.
17. Press RESET.
To reset transmiter to factory default settings:
1. Press PROGRAM, NEXT, NEXT, NEXT. The screen
should say "dEFAULt". Press ENTER.
2. Use the arrow keys to toggle between nO (retain your
configuration and calibration settings) and YES (restore
factory default settings to all variables).
3. Press ENTER, then EXIT.