MODEL 5081T-FF/FI INSTALLATION
INSTALLATION
UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the
shipper immediately for instructions. Save the box. If there
is no apparent damage, unpack the container. Be sure all
items shown on the packing list are present. If items are
missing, notify Rosemount Analytical immediately.
ROTATING THE DISPLAY
The Model 5081T-FF/F1 display can be rotated 90° left or
right. Disengage the cover lock and remove the front cover.
Remove the three screws holding the PCB stack and gen-
tly lift the display board. Do not disengage the ribbon cable
between the display board and the CPU board. Rotate the
display. The black infrared sensor will be at the top of the
display.
INSTALLATION
See Figure 4.
1. Although the analyzer is suitable for outdoor use, do not
install it in direct sunlight or in areas of extreme temper-
atures.
2. Install the analyzer in an area where vibrations and elec-
tromagnetic and radio frequency interference are mini-
mized or absent.
3. Keep the analyzer and sensor wiring at least one foot
from high voltage conductors. Be sure there is easy
access to the analyzer.
4. The conduit connections on the sides of the 5081 hous-
ing should be sealed to prevent moisture from entering
the enclosure.
5. The transmitter must not be mounted with both conduit
openings at the top
.
FIGURE 4. Mounting Model 5081T-FF/FI