80 Series V
Calibration Manual
14
Table 13. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator AC Voltage Range: 0 - 1000 V ac
Accuracy: ± 0.12 %
Frequency Range: 60 - 20000 Hz
Accuracy: ± 3 %
DC Voltage Range: 0 - 1000 V dc
Accuracy: ± 0.012 %
Current Range: 350 µA - 2 A
Accuracy: AC (60 Hz to 1 kHz): ± 0.25 %
DC: ± 0.05 %
Frequency Source: 19.999 kHz - 199.99 kHz
Accuracy: ± 0.0025 %
Amplitude: 150 mV to 6V rms
Accuracy: ± 5 %
Fluke 5520A Multi-Product
Calibrator or equivalent
Range: 1 Ω - 100 MΩ
Accuracy: 0.065 %
Function Generator Frequency = 900 kHz
Amplitude = 8.3V
Burst mode = 1
Burst rate = 100Hz
Burst Phase = -90 degrees
HP33120
Fluke 80 AK
TC Adapter Accessory
K-type Fluke 80 AK
K-type Thermocouple K-type, mini-plug on both ends
Performance Tests
The following performance tests verify the complete operability of the Meter and check
the accuracy of each Meter function against the Meter’s specifications. Performance tests
should be performed annually to ensure that the Meter is within accuracy specifications.
Accuracy specifications are valid for a period of one year after calibration adjustment,
when measured at an operating temperature of 18 °C to 28 °C and at a maximum of 90 %
relative humidity.
To perform the following tests, it is not necessary to open the case. No adjustments are
necessary. Make the required connections, apply the designated inputs, and determine if
the reading on the Meter display falls within the acceptable range indicated.
Note
If the Meter fails any of these tests, it needs calibration adjustment or repair.