Calibration Manual
18
Table 14. Accuracy Tests (cont.)
Display Reading
Step
Test
Function
Range 5500A Output
83 and 88 87
23 600 330 ( Use 2 wire Comp)
1
328.5 to 331.5
329.1 to 330.9
24 6 k 3.3 k (Use 2 wire Comp)
1
3.286 to 3.314
3.292 to 3.308
25 60 k 33 k
32.86 to 33.14
32.92 to 33.08
26 600 k 330 k
327.6 to 332.4
327.9 to 332.1
27 6 M 3.3 M
3.276 to 3.324
3.279 to 3.321
28
e
Ohms
50 M 30 M
29.67 to 30.33
29.67 to 30.33
29
60 nS Open input - 0.30 to 0.30 - 0.30 to 0.30
30
nS
Conductance
60 nS 100 M 9.60 to 10.40 9.60 to 10.40
31
G
Diode
6 V 3.0 V dc 2.939 to 3.061 2.939 to 3.061
32
?
AC Amps
6 A 3.0 A, 60 Hz 2.962 to 3.038 2.968 to 3.032
33
A
DC Amps
6 A 3.0 A 2.984 to 3.016 2.990 to 3.010
33B
A
DC Amps
10A
6
10A 9.94 to 10.06 9.96 to 10.04
34
60 mA 33 mA, 60 Hz 32.58 to 33.42 32.65 to 33.35
35
m
?
AC Milliamps
400 mA 330 mA, 60 Hz 325.8 to 334.2 326.5 to 333.5
36
60 mA 33 mA 32.83 to 33.17 32.89 to 33.11
37
m[
DC Milliamp
400 mA 330 mA 328.5 to 331.5 329.1 to 330.9
38
600 µA 330 µA, 60 Hz 325.8 to 334.2 326.5 to 333.5
39
U
?
AC
Microamps
6000 µA 3300 µA, 60 Hz 3258 to 3342 3265 to 3335
40
600 µA 330 µA 328.3 to 331.7 328.9 to 331.1
41
U[
DC
Microamps
6000 µA 3300 µA 3285 to 3315 3291 to 3309
42
10 nf Open input
2
0.21 to 0.31 0.21 to 0.31
43
100 nf 5 nf
5
04.7 to 05.3 04.7 to 05.3
44
E
Capacitance
100 µf 9.5 µf 09.2 to 09.8 09.2 to 09.8
45
1000 V 400 V, 400 Hz N/A 376 to 408
46
K
Low Pass
Filter
1000 V 400 V, 800 Hz
4
N/A 226 to 340
4