07303C DCN6646 9
2.0 SPECIFICATIONS
2.1 Specifications
Table 1 Specifications for the External Converter and the 501Y Bypass Pump
Converter Remotely mounted molybdenum external Converter with
temperature controller in a NEMA – 4 enclosure.
Converter Temperature
315
o
C 7
o
C with read-out on front panel of bypass pump chassis.
Tube/Electrical Cable
Assembly
Up to 50 feet (15 m) maximum length
Power, 501Y Bypass Pump
Assembly
120V~ 60 Hz, 230V~ 50 Hz 360 watts
Power, Converter Assembly 120V~ 50/60 Hz, 60 watts
Converter Efficiency >96% for NO2
Weight, 501Y Bypass Pump
Assembly
35 lbs (16 kg)
Weight, Remote Converter 30 lbs (11 kg)
501Y Bypass Pump Chassis
Dimensions HxWxD
9" x17" x23.6" (23cm x 43cm x 61cm)
Remote Converter Assy
Dimensions HxWxD
12"x19"x7" (31cm x 48cm x 18cm)
Environmental Installation Category (Over-voltage Category) II
Pollution Degree 2
Bypass Flow rate
Cal gas Flow rate
800 cc/min 10%, each channel
4000 cc/min (approx.)