18 MG302-EU-EN V2.0 4/11
Meg OHMS
Terminal Voltage Range Resolution Accuracy Test current Short circuit
current
0.125~4.000 MΩ 0.001MΩ +(2%+10)
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(4%+5)
125V (0%~+10%)
400.1~4000 MΩ 1MΩ +
(5%+5)
1mA @ load
125kΩ
≤1mA
0.250~4.000 MΩ 0.001MΩ +(2%+10)
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(3%+5)
250V (0%~+10%)
400.1~4000 MΩ 1MΩ +
(4%+5)
1mA @ load
250kΩ
≤1mA
0.500~4.000 MΩ 0.001MΩ +(2%+10)
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(2%+5)
500V (0%~+10%)
400.1~4000 MΩ 1MΩ +
(4%+5)
1mA @ load
500kΩ
≤1mA
1.000~4.000 MΩ 0.001MΩ +(3%+10)
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(2%+5)
1000V
(0%~+10%)
400.1~4000 MΩ 1MΩ +
(4%+5)
1mA @ load
1MΩ
≤1mA
Note: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.