19 MG302-EU-EN V2.0 4/11
Enclosure Double molded, waterproof (IP67)
Shock (Drop Test) 6.5 feet (2 meters)
Diode Test Test current 0.9mA maximum, open circuit voltage 2.8V DC typical
Storage capacity 8000 records
RF transmit distance 10 meters (approx)
Transmitter Frequency 433MHz
Continuity Check Audible signal will sound if the resistance is less than 35 (approx.), test
current <0.35mA
Peak Captures peaks >1ms
Temperature Sensor Requires type K thermocouple
Input Impedance >10MΩ VDC & >9MΩ VAC
AC Response True rms
ACV Bandwidth 50Hz to 1000Hz
Crest Factor ≤3 at full scale up to 500V, decreasing linearly to ≤1.5 at 1000V
Display 40,000 count, backlit, liquid-crystal display with bargraph
Overrange indication “OL” is displayed
Auto Power Off 15 minutes (approximately) with disable feature
Polarity Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate 2 times per second, nominal
Low Battery Indication “
” is displayed if battery voltage drops below operating voltage
Battery Six (6) 1.5V ‘AA’ batteries
Fuses mA, µA ranges; 0.5A/1000V ceramic fast blow (SIBA 70-172-40)
A range; 10A/1000V ceramic fast blow (SIBA 50-199-06)
Operating Temperature 5ºC to 40ºC (41ºF to 104ºF)
Storage Temperature -20
o
C to 60
o
C (-4
o
F to 140
o
F)
Operating Humidity Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC (104ºF)
Storage Humidity <80%
Operating Altitude 2000meters (7000ft) maximum
Weight 582g. (20.5 oz)
Size 200 x 92 x 50mm (7.8” x 3.6” x 1.9”)
Safety This meter is intended for origin of installation use and protected, against
the users, by double insulation per EN61010-1 and IEC61010-1 2
nd
Edition
(2001) to Category IV 600V and Category III 1000V; Pollution Degree 2.
Copyright © 2011 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com