Analyzer does not
function in the Ext.
Power switch position.
• Ensure the power
supply is plugged into
the External Power
receptacle or the
external battery is
plugged into the
External Battery
receptacle.
• Ensure there is power
to power supply.
Batteries do not charge.
• Ensure the power
supply is plugged into
the External Power
receptacle.
Batteries do not hold a
charge for more than a
few minutes.
• Replace batteries.
O
2
reading appears to be
incorrect.
• Recalibrate the sensor.
• Sensor may need to be
replaced.
O
2
reading does not
appear on the display in
the Program Menu.
• Cycle the SAMPLE
pump on and then off
then return to the
Program Menu.
Technical Support / Accessories
For technical assistance, please contact Teledyne Analytical Instruments.
Accessories & Recommended Spare Parts
• Carrying Case w/ shoulder strap
• External Battery Pack (includes carrying case)
• Oxygen Sensor (UFO-130)
• Replacement Filters
• Side Draw Needles
• Sample Probe Kit (needle, filter, tubing, and connector)
• Replacement Batteries
• Septum