Agilent Technologies E7405A TV Converter Box User Manual


 
Chapter 1 11
Making Basic Measurements
Comparing Signals
Figure 1-1 Placing a Marker on the 10 MHz Signal
9. Press Marker, Delta, to activate a second marker at the position of the
first marker.
10.Move the second marker to another signal peak using the
front-panel knob, or by pressing
Peak Search and then either
Next Pk Right or Next Pk Left. Next peak right is shown in Figure 1-2.
The amplitude and frequency difference between the markers is
displayed in the active function block and in the upper right corner
of the screen. See Figure 1-2.
11. The resolution of the marker readings can be increased by turning
on the frequency count function. For more information refer to
Making Better Frequency Measurements on page 22.
12.Press Marker, Off to turn the markers off.