Instruction Manual V1.2 - 11/2006
Instruments with sectioning motor:
The sectioning motor is now blocked.
All instruments:
The message ‘LOCKED’ in the display of control
panel 1 indicates that the handwheel has been
locked:
Unlocking:
• To unlock, push locking pin (2) to the left into
position (2a).
• Display indication ‘LOCKED’ disappears.
Instruments with sectioning motor:
The sectioning motor can now be activated
again.
2.2.2 Centering the handwheel grip
During motorized sectioning, for
safety reasons always center the
handwheel grip!
• To center grip (1), pull outwards and pivot
into center of handwheel.
• When released, the grip locks into position.
2.2.1 Locking the handwheel
Always cover the cutting edge with
the knife guard and lock the hand-
wheel:
• Prior to doing any work on
knife and/or specimen.
• Prior to exchanging specimens.
• During work breaks.
The handwheel can be locked in 2 positions:
• with the grip in the uppermost position (left),
• with the grip in the lowest position (right).
Locking:
• Rotate handwheel, until grip (1) is in upper or
lower position.
• To lock, press pin (2) to the right into position
(2b).
--> The upper locking position for pin (2) is
marked by a black dot (4).
2. Safety
2
2
CT-30°C OT-35°C
LOCKED