57
Leica CM 3050 S – Cryostat
Removal
• The lamp (3) is located behind the glare
shield (4).
• Carefully hold lamp with both of your hands.
Slightly pull the lamp downward at the left
side to remove it from fastening clip (5).
• Pull the lamp towards the left to remove it
from its socket (6).
Installing the replacement lamp
• Hold the lamp in correct position (see draw-
ing below) and push towards the right until it
locks into the socket.
• Push the lamp slightly upwards until it locks
into the clip.
• Put the sliding window back in place.
• Reconnect the instrument to mains and
switch it back on.
7.6 Exchanging the fluorescent light lamp
Prior to exchanging the lamp,
please read and observe the fol-
lowing safety instructions:
• ‘Maintenance’
(Chapter 2.3.13)
Removing the sliding window
• Hold sliding window (1) at grip bar (2), lift
slightly and pull out towards yourself.
7. Disinfection, cleaning and maintenance