3.4 Installing/inserting accessories
3.4.1 Installing the handwheel
Insert pin (1) of the handwheel shaft into
hole (2).
Place spring washer (3) onto screw (4) as
shown.
Tighten screw (4) with an Allen key.
Cover with selfadhesive disc (5).
For purposes of transport (e.g. narrow
doors), the handwheel can be
removed.
To remove the handwheel, proceed as de-
scribed above but in reverse order.
3. Installation
3.4.2 Inserting the accessories
Place the rubber mat on top of the housing.
Insert the storage shelves into the cryo-
chamber.
Install the stationary heat extractor into the
quick-freeze shelf (see also drawing in
chapter 5.5.1 ‘Freezing specimens onto
specimen discs’).
Insert the low temperature stabilizer into the
quick freeze shelf (it must be located in the
pivoting range of the heat extractor - see
also drawing in chapter 5.5.1 ‘Freezing spec-
imens onto specimen discs’).
Insert section waste tray and brush shelf.
Install knife holder base onto microtome
base plate and clamp.
Install knife holder and clamp (see knife
holder instruction manual for details).
Place knife case with knife into chamber to
precool.
Place all tools needed for section prepara-
tion into the chamber.
Close the sliding window.
For a complete overview of all
individual parts:
--> see Chapter 4.1 ‘Overview’.