Leica CM3050 S Projection Television User Manual


 
28
Instruction Manual V1.2 - 11/2006
5. Operation
Meaning of indication:
Operator activity:
yes
no
no
Acknowledge indication by
pressing
once briefly.
Initialization phase 1
Initialization phase 2
Software version
Selected display language
Sectioning motor speed
level ‘slow’ / (fast)
---
---
---
Select language via
keys
Select speed level via
keys
5.3.3 Display indications when switching on
Standard indication ‘locked’ is displayed:
(Display background dark)
To unlock: press
until the display back-
ground lights up:
yes
Temperature prior to
switching off < 0 °C
If this indication is displayed: Do not switch on
the instrument!
Prior to switching on, both microtome and
accessories must be completely dry!
Remove microtome/accessories, disinfect if
applicable, then dry thoroughly and reinstall.
(--> See chapters 7.1 - 7.5).
*) Indication in instruments with specimen cooling. Instruments without specimen cooling: only CT (centered) is displayed.
Temperature upon
switching on 0 °C
8888888888888888
8888888888888888


CM3 0 5 0S Vs . XX . XX
ENGL I SH
CUTTING MOTOR
SLOW
POWER FA I LURE
12:00
CT....°C OT....°C

HOME ❄❄
CT. . ..°C OT . . . . ° C *)
DRY MICROTOME
CT....°C OT....° C *)
CT. . ..°C OT . . . . ° C *)