Omega M-4292/1006 Universal Remote User Manual


 
Integer
For most computers and controllers they are decimal numbers
ranging from -32768 to 32768. For example:
FAULT? could return 110
Refer to the Error Codes table (Table 8) for more information on error
codes.
Floating
Floating numbers have up to 15 significant figures and exponents.
For example:
ZERO_MEAS returns 1.210000E
-
2, PSI
Character Response Data (CRD)
Data returned as keywords. For example:
PRES_UNIT? returns PSI, NONE, NONE
Indefinite ASCII (IAD)
Any ASCII characters followed by a terminator. For example:
*IDN? returns BETA, 301, 250, 1.00
9.4.4 Calibrator Status
Error Queue
If an error occurs due to invalid input or buffer overflow, its error code
is sent to the error queue. The error code can be read from the
queue with the command FAULT?. The error queue holds 15 error
codes. When it is empty, FAULT? returns 0. The error queue is
cleared when power is reset or when the clear command *CLS is
entered.
Input Buffer
Calibrator stores all received data in the input buffer. The buffer holds
250 characters. The characters are processed on a first in, first out
basis.
26