FUNC
Sets the display indicated in argument one to the function
indicated in argument 2.
For example:
To set the lower display to P1 Pressure mode send FUNC
LOWER,P1.
FUNC?
Returns the current mode of all displays. For example if the
calibrator is set to [P1] ST on the upper display, [P1] on the
middle, and the lower display is off, FUNC? Would return:
ST_P1,P1,NONE
HI_ERR
Sets the 100% point for the percent error mode calculation in the
current engineering units.
For example:
To set the 100% point to 100 psi send HI_ERR 100.
HI_ERR?
Returns the 100% point for the percent error mode calculation.
For example:
If the 100% point is set to 100 psi, HI_ERR? would return
1.000000E+02, PSI .
LOCAL
Restores the calibrator to local operation if it was in remote
mode. Also clears LOCKOUT if the calibrator was in lockout
mode.
LOCKOUT
Sending this command sets the lockout state, when the unit is in
REMOTE or goes to remote it prohibits use of the keypad
completely. The lockout state can only be cleared by sending
the LOCAL command.
32