Omega M-4292/1006 Universal Remote User Manual


 
29
Table 8: Error Codes
Error Number Error Description
100 A non-numeric entry was received where it should be a
numeric entry
101 Too many digits entered
102 Invalid units or parameter value received
103 Entry is above the upper limit of the allowable range
104 Entry is below the lower limit of the allowable range
105 A required command parameter was missing
106 An invalid command parameter was received
107 Pressure not selected
108 Invalid sensor type
109 Pressure module not connected
110 An unknown command was received
111 Bad Parameter received
112 The serial input buffer overflowed
113 Too many entries in the command line
114 The serial output buffer overflowed
9.6 Entering Commands
Commands for the calibrator may be entered in upper or lower case.
There is at least one space required between the command and
parameter, all other spaces are optional. Almost all commands for
the calibrator are sequential; any overlapped commands will be
indicated as such. This section will briefly explain each of the
commands and describe their general use, which will include any
parameters that may be entered with the command as well as what
the output of the command is.
9.6.1 Common Commands
*CLS
Clears the error queue. Also terminates all pending operations. When
writing programs, use before each procedure to avoid buffer
overflow.
*IDN?
Returns the manufacturer, model number, and firmware revision of
the Calibrator. For example:
*IDN? will return BETA, 301, 0, 1.00