One-Button Measurement Functions
Meas Setup (Burst Power)
One-Button Measurement Functions
points.
RMS — Selects averaging that converts trace data from dB to power units, then
averages the power trace data. This selection requires more time to perform.
Key Path:
Meas Setup
Factory Preset: RMS
Range: RMS or Log
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
[:SENSe]:BPOWer:AVERage:TYPe LPOWer|POWer
[:SENSe]:BPOWer:AVERage:TYPe?
Example: BPOW:AVG:TYP LPOWer to select Log type.
BPOW:AVG:TYP?
2.5.4 Threshold Lvl
Enables you to set the level above which the mean carrier power calculation is based. The
threshold level can be described in dB (
Rel) or dBm (Abs).
Key Path:
Meas Setup
Factory Preset: 30 dB
Terminators: dB or dBm
Default Terminator dB
Knob Increment: 0.1 dB/dBm
Step Key
Increment: 6 dB/dBm
Range: –60 dBm to 60 dBm (in absolute mode)
–60 dB to 0 dB (in relative mode)
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
[:SENSe]:BPOWer:THReshold <number>
[:SENSe]:BPOWer:THReshold?
[:SENSe]:BPOWer:THReshold:TYPE ABSolute|RELative