Agilent Technologies E4443A(3HZ-6.7GHZ) TV Converter Box User Manual


 
Chapter 2 129
One-Button Measurement Functions
Marker (Complementary Cumulative Distribution Function—CCDF)
One-Button Measurement Functions
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X:POSition?
This defines marker X position in display points from 0 to 50 dB.
Example: CALC:PST:MARK:X 30 dB activates a normal marker (marker 1) at the 30
dB location on the specified trace (see “Marker Trace” on page 127).
CALC:PST:MARK2:X:POS 10 selects marker 2 and moves it to 1 dB location
on the specified trace.
2.9.8 Marker Y Position (Remote Command Only)
Displays the marker Y position (vertical) of the specified trace (see “Marker Trace” on
page 127) in the current Y-axis units (percent).
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4:Y?
Queries the marker Y value or delta in the current y axis units. The marker must be
ON for
the response to be valid.
Example: CALC:PST:MARK1:Y?
2.9.9 Marker Maximum and Minimum (Remote Command Only)
Moves the marker to the maximum or minimum detected amplitude value on the display.
If the marker is off, this command has no effect.
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MAXimum
:CALCulate:PSTatistic:MARKer[1]|2|3|4:MINimum
Example: CALC:PST:MARK:MAX
CALC:PST:MARK3:MIN