Agilent Technologies E4443A(3HZ-6.7GHZ) TV Converter Box User Manual


 
66 Chapter 2
One-Button Measurement Functions
MEASURE (Spectrum Analysis Mode)
One-Button Measurement Functions
Remote Command Notes: The main channel power is returned in the current amplitude
units, and the lower and upper channel results are always returned in dB.
Example: FETC:ACP? or MEAS:ACP? or READ:ACP? commands return the scalar results
of main channel power, lower channel power (relative), and upper channel
power (relative) if only one offset is set to on and the Radio Standard is set
to None. Otherwise, 28 values are returned. If
Meas Type is Total Power Ref,
these are the main channel power in dBm and in dBc from the carrier (0
dBc) each repeated, followed by the absolute and relative power levels for
each lower and upper offset if Meas Type is Total Power Ref. Otherwise
they are the main channel power spectral density and relative power
spectral density (0 dB) each repeated. Then followed by the absolute and
relative spectral density values for each lower and upper offset.