One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
2.3.7 Method
Enables you to set the measurement method.The resolution bandwidth method is most
useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth method
is preferred for other signals.
NOTE ESA Only:
This key toggles to select either the integration bandwidth method (IBW) or
the resolution bandwidth method (RBW).
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: IBW
History: Added with PSA firmware revision A.02.00.
Added with ESA firmware revision A.08.00.
Remote Command:
[:SENSe]:ACPower:METHod IBW|RBW|FAST
[:SENSe]:ACPower:METHod?
Remote Command Notes:
ESA Only: Fast is not an available option.
Example: ACP:METH RBW
ACP:METH?
2.3.7.1 IBW
PSA Only: Sets the measurement control method to IBW.
Key Type: 1 of N menu
Key Path:
Meas Setup, Method
History: Modified with PSA firmware revision A.06.00
Remote Command:
Use[:SENSe]:ACPower:METHod See “Method” on page 97.
Example: ACP:METH IBW