Hitachi AP 720STM Cable Box User Manual


 
19
Figure 39 - Alarm Screen of No Pettes
[Screen Display]
(Figure 39)
[Summary]
If the control software detects that there is no pette at the start of the run, Figure 39
will appear. If the run needs to be continued, setup pettes in the Pette Rack, and
press ‘Retry’ key to continue. If the run needs to be cancelled, press ‘End’ key to
end the Auto-Run.
Figure 40 - Alarm Screen of Wrong Pette Setup
[Screen Display]
(Figure 40)
[Summary]
If the control software detects that the pette setup is wrong at the start of the run,
Figure 40 will appear. If the run needs to be continued, setup pettes correctly and
press ‘Retry’ key to continue. If the run needs to be cancelled, press ‘End’ key to
end the Auto-Run.
Figure 40 shows that the last pette is detected at position #10, and a wrong pette
setup is detected at position #12. This is because no pette is detected at position
#11.
When “End # 0” is shown, pette is not detected at position #1.