Omega RD2800 DVD Recorder User Manual


 
-92-
14. PRINTING FORMAT
14.1 Programming Automatic Range-Shift Printing
This programming display only appears for the
printing format (option). This programming is to
change the trace printing range automatically.
1 Automatic Range-shift
The automatic range-shift function changes
the trace printing range up to 5 stages
according to the measured values.
(1) This programming can be applied to each
channel.
(2) The total printing range can be
programmed optionally irrespective of the
programmed range in the [Range/Printing
range].
(3) For each range, refer to the minimum
printing range in section 22.1 (Input
Specifications). The printing may be
dispersed if each range is programmed to
be less than the minimum printing range.
(4) The range shifting is executed when the
measured value exceeds approx. 0.5 mm
from the minimum (zero) or the maximum
(span) range.
2 Programming Mode
1) Screen 1 (Printing format check)

 

Dots appear at the selected format.
By pressing , the dots disappear and the
cursor appears instead.
2) Screen 2 (Programming channel)

By selecting automatic range-shift (

), the cursor
appears at the channel programming position.
3) Screen 3 (Zero programming at No.1 range)
 
(1) The default programmed value is the minimum
value in the [Range/Printing range]. This
programming can be changed as required.
(2) When all programming are completed, return to
the above display and execute [Storing].
4) Screen 4 (Span programming at No. 1 to 5
ranges)
 


(1) When the programming a span, the range
number advances. Program the span up to the
required range number.
(2) After pressing up to range No.5, press
again to return to Display 3.
Mark is the programming point and No. (1) to (5) shows
the order of programming. By pressing when the
programming ends, the screen moves to the next
programming screen. As No. 5 is not used in this example,
only press .
Printing range
for 1st range
-100
400
(1)
(2)
Printing range
for 2nd range
(3)
(400)
600
Printing range
for 3rd range
(4)
(600)
800
Printing range
for 4th range
(5)
(800)
1200
-100
400
600
800 1200
1st range
(R1)
4th range
(R4)
2
n
d
range
(R2)
3
r
d
range
(R3)
-200 1370
Measuring value
<Example> Automatic range-shift programming
<Range No. 23 (K thermocouple)>
Standard
Automatic
range-shift
Zone
Compressed/expanded
Programming value
(Max. 5 digits by left-justify)
Range No.
Span
T
ota
l
pr
i
nt
i
ng
range
ENTRY
ENTRY
SET
END
ENTRY
ENTRY
A
utomatic range selection
Channel
Cursor
Programming value
(Max. 5 digits by left-justify)
Range No.
Zero