3
4. The operator should take precautions to avoid
touching the rear panel input and output circuitry and
the patient at the same time.
5. Model LMD-1950MD is a monitor intended for use
in a medical environment to display pictures from
cameras or other systems. These models are
nonpatient care equipment with respect to the
requirement of leakage current.
FOR CUSTOMERS IN EUROPE
This equipment has been found to comply with limits for
a Class B device pursuant to EN60601-1-2. However, if
this equipment does cause harmful interference to other
devices, which can be determined by turning this
equipment off and on, the user is encouraged to try to
correct the interference by one or more of the following
measures:
• Increase the separation between the equipment and
other devices.
• Connect the equipment into an outlet on a circuit
different from that to which other devices are
connected.
• Consult the dealer or an experienced radio/TV
technician for help.
Caution
When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.
Symbols on the unit
Symbol Location This symbol indicates
Bottom Power switch.
Press to turn the monitor
on.
Bottom Power switch.
Press to turn the monitor
off.
Bottom The equipotential
terminal which brings the
various parts of a system
to the same potential.
Bottom Functional earth terminal
8
Front Brightness control
6
Front Contrast control
-
Front Key inhibit
The setting are locked so
that they cannot be
changed.
Warning on power connection
Use a proper power cord for your local power supply.
The United States Canada Continental Europe
Plug type HOSPITAL GRADE
E41395
E41395-A
HOSPITAL GRADE
LL33182
LL76662
LP-34A
Female end LS-60
Cord type H05VV-F
Rated Voltage & Current 10A/125V 10A/125V 10A/250V
Safety approval UL CSA VDE