Agilent Technologies N8201A TV Converter Box User Manual


 
46 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
2 Hardware Setup and Configuration
Measurement Procedure
5 Tune the Spectrum Analyzer to the following:
Frequency: 7.5 MHz
Amplitude: 10 dBm
Span: 100 kHz
6 Tune the downconverter to the following:
Level: Attenuator: 20 dBm
RF Enabled: Enabled
Input RF: Signal Source: 50 MHz Cal
Frequency: 50 MHz
7 Verify that a 7.5 MHz signal is present on the spectrum analyzer.