Agilent Technologies N8201A TV Converter Box User Manual


 
Using the Agilent Synthetic Instrument GUI 3
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz 79
Settling Mode
Selects the operating mode for the Settling Channel: Disabled, Driven, or Wired OR.
Wired OR is only available for LXI channels.Use Wired OR if more than one instrument is
driving the LXI channel. When only one instrument is driving the LXI channel then use the
Driven mode.
Settling Destination Path
The target of the LAN event. The Settling Destination Path can be either an IP address, a
hostname, or 'All'. (The function is only visible if Settling Channel is a LAN channel and can
be set if the Settling mode is not Disabled.)
Settling Slope
Determines whether an event will occur on positive or negative edges. Possible values are:
Positive or Negative.
Execute Settling Event
Starts a settling event after all of the parameters are define.
Event Logging
When enabled, a history of all instrument errors and messages are displayed.
Enable Logging
Enables or disables event logging.
Event Log
Displays the event log in the right pane when the Event Log page is selected.
Clear Event Log
Clears the event log.
Disable All Events
Prevents any previously events from occurring.
NOTE
Settling mode must be enabled for Settling Slope to be set.