Agilent Technologies N8201A TV Converter Box User Manual


 
Hardware Setup and Configuration 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz 49
Measurement Procedure
4 Tune the Spectrum Analyzer to the following:
Frequency: 7.5 MHz
Amplitude: 10 dBm
Span: 100 kHz
5 Tune the downconverter to the following:
Center Frequency: 3.5 GHz
Attenuator: 20 dBm
RF Enabled: Enabled
Source: Internal
6 Tune the Source to the following:
Frequency: 3.5 GHz
Amplitude: 0 dBm
RF On
Verify that a 7.5 MHz signal is present on the spectrum analyzer.