Hardware Setup and Configuration 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz 49
Measurement Procedure
4 Tune the Spectrum Analyzer to the following:
• Frequency: 7.5 MHz
• Amplitude: 10 dBm
• Span: 100 kHz
5 Tune the downconverter to the following:
• Center Frequency: 3.5 GHz
• Attenuator: 20 dBm
• RF Enabled: Enabled
• Source: Internal
6 Tune the Source to the following:
• Frequency: 3.5 GHz
• Amplitude: 0 dBm
• RF On
Verify that a 7.5 MHz signal is present on the spectrum analyzer.